  5. Gas Powered Emergency Resuscitator Recall: Mercury Medical Removes Neo-Tee T-Piece Resuscitator Due to An Undersized Spring in The Controller That May Prevent the Device from Delivering the Required Pressure Levels Needed for Effective Ventilation
  1. Medical Device Recalls

Gas Powered Emergency Resuscitator Recall: Mercury Medical Removes Neo-Tee T-Piece Resuscitator Due to An Undersized Spring in The Controller That May Prevent the Device from Delivering the Required Pressure Levels Needed for Effective Ventilation

This recall involves removing certain devices from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it. 

Affected Product

Figure of Non In-line and In-line versions of the Neo-Tee Resuscitator.

What to Do

On December 5, 2024, Mercury Medical sent all affected customers an Urgent Field Safety Notice recommending the following actions:

  • Immediately check the inventory for affected product, within the scope of this recall
  • Stop use and distribution and immediately quarantine the affected product.
  • Provide a copy of this recall notice to all customers who have received impacted product. Notify all personnel subject to using these devices.

Reason for Recall

Mercury Enterprises, Inc. dba Mercury Medical is recalling the Neo-Tee T-Piece resuscitator due to a small spring in the controller that may prevent the device from delivering the required pressure levels needed for effective ventilation. This issue could reduce positive pressure, affecting the patient's breathing support.

The use of affected product may cause serious adverse health consequences, including low oxygen levels (desaturation), slow heart rate (bradycardia), lack of oxygen (hypoxia), high carbon dioxide levels (hypercarbia), and death.

There have been no reported injuries or death associated with this issue.

Device Use

The Neo-Tee T-Piece Resuscitator is a gas-powered emergency resuscitator intended to provide emergency respiratory support by means of a facemask or a tube inserted into a patient's airway. It is intended for use with pediatric patients weighing less than 10kg (22lbs). The device is designed with in-line flow controller, in the circuit.

Contact Information

Customers in the U.S. with questions about this recall should contact Mercury Medical at 1-727-573-0088. 

Full List of Affected Devices

Part NumberUnique Device IDLot Number(s)
105080510641043508053, 306410435080572421450805, 2423550805, 2423650805, 2426750805
105080810641043508084, 306410435080882426450808, 2426750808, 2426850808, 2429050808, 2429350808
105080910641043508091, 306410435080952426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809
105081010641043508107, 306410435081012426750810, 2429350810, 2429750810, 2429850810, 2430250810
105081110641043508114, 306410435081182426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811
105081410641043508145, 306410435081492426450814, 2426750814, 2429050814, 2429750814, 2430250814,
105083210641043508329, 306410435083232429050832, 242985083
105083910641043508398, 306410435083922430350839
105084010641043508404, 306410435084082429050840, 2429750840
105084110641043508411, 306410435084152429050841
105084210641043508428, 306410435084222426750842

Additional FDA Resources

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.