58.81 (a) A testing facility shall have standard operating procedures in writing setting forth nonclinical laboratory study methods that management is satisfied are adequate to insure the quality and integrity of the data generated in the course of a study. |
792.81 (a) A testing facility shall have standard operating procedures in writing, setting forth study methods that management is satisfied are adequate to insure the quality and integrity of the data generated in the course of a study. |
Section II 7.1. A test facility should have written Standard Operating Procedures approved by test facility management that are intended to ensure the quality and integrity of the data generated by that test facility. |
Deviations from SOPs |
58.81 (a) All deviations in a study from standard operating procedures shall be authorized by the study director and shall be documented in the raw data. |
792.81 (a) All deviations in a study from standard operating procedures shall be authorized by the study director and shall be documented in the raw data. |
Section II 7.3. Deviations from Standard Operating Procedures related to the study should be documented and should be acknowledged by the Study Director and the Principal Investigator(s), as applicable. |
Changes in SOPs |
58.81 (a) Significant changes in established standard operating procedures shall be properly authorized in writing by management. |
792.81 (a) Significant changes in established standard operating procedures shall be properly authorized in writing by management. |
Section II 7.1. Revisions to Standard Operating Procedures should be approved by test facility management. |
Required SOPs |
58.81
|
792.81
|
Section II
|
Availability of SOPs |
58.81 (c) Each laboratory area shall have immediately available laboratory manuals and standard operating procedures relative to the laboratory procedures being performed. Published literature may be used as a supplement to standard operating procedures. |
792.81 (c) Each laboratory or other study area shall have immediately available manuals and standard operating procedures relative to the laboratory or field procedures being performed. Published literature may be used as a supplement to standard operating procedures. |
Section II 7.2. Each separate test facility unit or area should have immediately available current Standard Operating Procedures relevant to the activities being performed therein. Published text books, analytical methods, articles and manuals may be used as supplements to these Standard Operating Procedures. |
Historical file of SOPs |
58.81 (d) A historical file of standard operating procedures, and all revisions thereof, including the dates of such revisions, shall be maintained. |
792.81 (d) A historical file of standard operating procedures, and all revisions thereof, including the dates of such revisions, shall be maintained. |
Section II 10.1. (f) The historical file of all Standard Operating Procedures; |
Labeling of Reagents |
58.83 All reagents and solutions in the laboratory areas shall be labeled to indicate identity, titer or concentration, storage requirements, and expiration date. Deteriorated or outdated reagents and solutions shall not be used. |
792.83 All reagents and solutions in the laboratory areas shall be labeled to indicate identity, titer or concentration, storage requirements, and expiration date. Deteriorated or outdated reagents and solutions shall not be used. |
Section II 4.4. Chemicals, reagents, and solutions should be labelled to indicate identity (with concentration if appropriate), expiry date and specific storage instructions. Information concerning source, preparation date and stability should be available. The expiry date may be extended on the basis of documented evaluation or analysis. |
Records of Test System Receipt |
Section II 5.2.3. Records of source, date of arrival, and arrival condition of test systems should be maintained. |
||
Test System Care SOPs |
58.90 (a) There shall be standard operating procedures for the housing, feeding, handling, and care of animals. |
792.90 (a) There shall be standard operating procedures for the housing, feeding, handling, and care of animals and other test systems. |
Section II 5.2.1. Proper conditions should be established and maintained for the storage, housing, handling and care of biological test systems, in order to ensure the quality of the data. |
Isolation of New Test Systems |
58.90 (b) All newly received animals from outside sources shall be isolated and their health status shall be evaluated in accordance with acceptable veterinary medical practice. |
792.90 (b) All newly received test systems from outside sources shall be isolated and their health status or appropriateness for the study shall be evaluated. This evaluation shall be in accordance with acceptable veterinary medical practice or scientific methods. |
Section II 5.2.2. Newly received animal and plant test systems should be isolated until their health status has been evaluated. If any unusual mortality or morbidity occurs, this lot should not be used in studies and, when appropriate, should be humanely destroyed. At the experimental starting date of a study, test systems should be free of any disease or condition that might interfere with the purpose or conduct of the study. Test systems that become diseased or injured during the course of a study should be isolated and treated, if necessary to maintain the integrity of the study. Any diagnosis and treatment of any disease before or during a study should be recorded. |
Test System Disease Conditions |
58.90 (c) At the initiation of a nonclinical laboratory study, animals shall be free of any disease or condition that might interfere with the purpose or conduct of the study. If, during the course of the study, the animals contract such a disease or condition, the diseased animals shall be isolated, if necessary. These animals may be treated for disease or signs of disease provided that such treatment does not interfere with the study. The diagnosis, authorizations of treatment, description of treatment, and each date of treatment shall be documented and shall be retained. |
792.90 (c) At the initiation of a study, test systems shall be free of any disease or condition that might interfere with the purpose or conduct of the study. If during the course of the study, the test systems contract such a disease or condition, the diseased test systems should be isolated, if necessary. These test systems may be treated for disease or signs of disease provided that such treatment does not interfere with the study. The diagnosis, authorization of treatment, description of treatment, and each date of treatment shall be documented and shall be retained. |
Section II 5.2.2. At the experimental starting date of a study, test systems should be free of any disease or condition that might interfere with the purpose or conduct of the study. Test systems that become diseased or injured during the course of a study should be isolated and treated, if necessary to maintain the integrity of the study. Any diagnosis and treatment of any disease before or during a study should be recorded. |
Test System Identifcation |
58.90 (d) Warm-blooded animals, excluding suckling rodents, used in laboratory procedures that require manipulations and observations over an extended period of time or in studies that require the animals to be removed from and returned to their home cages for any reason (e.g., cage cleaning, treatment, etc.), shall receive appropriate identification. All information needed to specifically identify each animal within an animal-housing unit shall appear on the outside of that unit |
792.90 (d) Warm-blooded animals, adult reptiles, and adult terrestrial amphibians used in laboratory procedures that require manipulations and observations over an extended period of time, or in studies that require these test systems to be removed from and returned to their test system-housing units for any reason (e.g., cage cleaning, treatment, etc.), shall receive appropriate identification (e.g., tattoo, color code, ear tag, ear punch, etc.). All information needed to specifically identify each test system within the test system-housing unit shall appear on the outside of that unit. Suckling mammals and juvenile birds are excluded from the requirement of individual identification unless otherwise specified in the protocol. |
Section II 5.2.5. All information needed to properly identify the test systems should appear on their housing or containers. Individual test systems that are to be removed from their housing or containers during the conduct of the study should bear appropriate identification, wherever possible. |
Housing of Test Systems |
58.90 (e) Animals of different species shall be housed in separate rooms when necessary. Animals of the same species, but used in different studies, should not ordinarily be housed in the same room when inadvertent exposure to control or test articles or animal mixup could affect the outcome of either study. If such mixed housing is necessary, adequate differentiation by space and identification shall be made. |
792.90
|
Section II 5.2.1. Proper conditions should be established and maintained for the storage, housing, handling and care of biological test systems, in order to ensure the quality of the data. 5.2.7. Test systems used in field studies should be located so as to avoid interference in the study from spray drift and from past usage of pesticides. |
Cleaning of Cages & Equipment |
58.90 (f) Animal cages, racks and accessory equipment shall be cleaned and sanitized at appropriate intervals. |
792.90 (f) Cages, racks, pens, enclosures, aquaria, holding tanks, ponds, growth chambers, and other holding, rearing, and breeding areas, and accessory equipment, shall be cleaned and sanitized at appropriate intervals. |
Section II 5.2.6. During use, housing or containers for test systems should be cleaned and sanitised at appropriate intervals. Any material that comes into contact with the test system should be free of contaminants at levels that would interfere with the study. Bedding for animals should be changed as required by sound husbandry practice. Use of pest control agents should be documented. |
Animal Feed & Water |
58.90 (g) Feed and water used for the animals shall be analyzed periodically to ensure that contaminants known to be capable of interfering with the study and reasonably expected to be present in such feed or water are not present at levels above those specified in the protocol. Documentation of such analyses shall be maintained as raw data. |
792.90 (g) Feed, soil, and water used for the test systems shall be analyzed periodically to ensure that contaminants known to be capable of interfering with the study and reasonably expected to be present in such feed, soil, or water are not present at levels above those specified in the protocol. Documentation of such analyses shall be maintained as raw data. |
Section II 5.2.6. Any material that comes into contact with the test system should be free of contaminants at levels that would interfere with the study. |
Animal Bedding |
58.90 (h) Bedding used in animal cages or pens shall not interfere with the purpose or conduct of the study and shall be changed as often as necessary to keep the animals dry and clean. |
792.90 (h) Bedding used in animal cages or pens shall not interfere with the purpose or conduct of the study and shall be changed as often as necessary to keep the animals dry and clean. |
Section II 5.2.6. Any material that comes into contact with the test system should be free of contaminants at levels that would interfere with the study. Bedding for animals should be changed as required by sound husbandry practice. |
Pest Control Program |
58.90 (i) If any pest control materials are used, the use shall be documented. Cleaning and pest control materials that interfere with the study shall not be used. |
792.90 (i) If any pest control materials are used, the use shall be documented. Cleaning and pest control materials that interfere with the study shall not be used. |
Section II 5.2.6. Use of pest control agents should be documented. |
Acclimation of Test System |
792.90 (j) All plant and animal test systems shall be acclimatized to the environmental conditions of the test, prior to their use in a study. |
Section II 5.2.4. Biological test systems should be acclimatised to the test environment for an adequate period before the first administration/application of the test or reference item. |