Notification Protocols When a Facility Moves a Mammography Unit
The following tables are based on information supplied by the MQSA accreditation bodies and State certifying agencies and are to be used to advise a facility when it moves a mammography x-ray unit. This is a general outline and by its very nature cannot address all aspects of a specific move. The facility may contact its accreditation body or certifying agency if it has additional questions.
Scenario 1. A mammography x-ray unit that is on wheels and can be plugged into an ordinary electrical outlet is moved to another room within the same building. The guidance for this Scenario does NOT apply to the situation where the unit is in a truck and is wheeled off to do patients.
Action | ABs | CAs | |||||||
---|---|---|---|---|---|---|---|---|---|
ACR | AR | IA | TX | FDA | IL | IA | SC | TX | |
Notify AB/CA prior to move | No | No | No | No | No | No | No | Yes | No |
Notify AB/CA within 30 days of move | No | Yes | Yes | No | No | No | Yes | Yes | No |
Post-Move Pre-Exam test* required | Yes ACR also recommends checking with MP | Yes | Yes | Yes | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes - also recommend checking with MP |
MEE required | No | No | No | No | No | No | No | No | No |
Change in accreditation/ certification status | No | No | No | No | No | No | No | No | No |
FDA Form 2579 needs to be filed by the installer** | NA | NA | NA | NA | No | NA | NA | NA | NA |
*The post-move pre-examination test is described in the Policy Guidance Help System under “Quality Assurance/Equipment/Mobile Units Equipment Quality Control - Question #3.” This post-move pre-examination testing does not have to be performed by a medical physicist.
**The filing of Form 2579 is NOT an MQSA requirement. Under the Diagnostic X-Ray Performance Standards (21 CFR 1020.30(d)(1)), Form 2579 needs to be filed by the installer with the FDA.
Scenario 2. A mammography x-ray unit not meeting the criteria in example #1 is moved to another room within the same building.
Action | ABs | CAs | |||||||
---|---|---|---|---|---|---|---|---|---|
ACR | AR | IA | TX | FDA | IL | IA | SC | TX | |
Notify AB/CA prior to move | No | No | Yes | Yes | No | Yes | Yes | Yes | No |
Notify AB/CA within 30 days of move | No | Yes | No | No | No | No | No | Yes | No |
Post-Move Pre-Exam test* required | Yes*** | No | Check with accreditation body | No | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | No |
MEE required | No*** | Yes | Check with accreditation body | Yes | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Yes, checked at time of inspection |
Change in accreditation/ certification status | No | No | No | No | No | No | No | No | No |
FDA Form 2579 needs to be filed by the installer** | NA | NA | NA | NA | Yes | NA | NA | NA | NA |
*The post-move pre-examination test is described in the Policy Guidance Help System under “Quality Assurance/Equipment/Mobile Units Equipment Quality Control - Question #3.” This post-move pre-examination testing does not have to be performed by a medical physicist.
** The filing of Form 2579 is NOT an MQSA requirement. Under the Diagnostic X-Ray Performance Standards (21 CFR 1020.30(d)(1)), Form 2579 needs to be filed by the installer with the FDA.
***The facility needs to check with its medical physicist to determine if disassembly/reassembly has occurred. If it has, the Post-Move Pre-Exam test is to be replaced by an MEE performed by the medical physicist prior to using the unit on patients. The MEE does not have to be sent to the ACR.
Scenario 3. A mammography x-ray unit is moved to another building.
Action | ABs | CAs | |||||||
---|---|---|---|---|---|---|---|---|---|
ACR | AR | IA | TX | FDA | IL | IA | SC | TX | |
Notify AB/CA prior to move | Yes Call ACR for instructions | No | Yes | Yes | No | Yes | Yes | Yes | Yes |
Notify AB/CA within 30 days of move | No | Yes | No | No | No | No | No | Yes | No |
Post-Move Pre-Exam test* required | Possible Follow ACR’s instructions from initial call above | No | Check with accreditation body | No | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | No |
MEE required | Possible Follow ACR’s instructions from initial call above | Yes | Check with accreditation body | Yes | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Yes, checked at time of inspection |
Change in accreditation/ certification status | Possible Status will be determined during initial call | No | Possible | Possible | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB | Same as facility’s AB |
FDA Form 2579 needs to be filed by the installer** | NA | NA | NA | NA | Yes | NA | NA | NA | NA |
*The post-move pre-examination test is described in the Policy Guidance Help System under “Quality Assurance/Equipment/Mobile Units Equipment Quality Control - Question #3.” This post-move pre-examination testing does not have to be performed by a medical physicist.
** The filing of Form 2579 is NOT an MQSA requirement. Under the Diagnostic X-Ray Performance Standards (21 CFR 1020.30(d)(1)), Form 2579 needs to be filed by the installer with the FDA.
Scenario 4. A mammography x-ray unit is moved to a different facility (different MQSA ID number), even if the unit is on wheels and can be plugged into an ordinary electrical outlet.
Action | ABs | CAs | |||||||
---|---|---|---|---|---|---|---|---|---|
ACR | AR | IA | TX | FDA | IL | IA | SC | TX | |
Notify AB/CA prior to move | Yes | Yes | Yes | Yes | No | Yes | Yes | Yes | Yes |
Notify AB/CA within 30 days of move | No | Yes | No | No | No | No | No | Yes | No |
Post-Move Pre-Exam test* required | No | No | No | No | No | No | No | No | No |
MEE required | Yes | Yes | Yes | Yes | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes, checked at time of inspection | Yes, checked at time of inspection |
Change in accreditation/ certification status | Yes Unit must go through new accreditation process | Yes Unit must go through new accreditation process | Yes Unit must go through new accreditation process | Yes Unit must go through new accreditation process | Yes | Yes | Yes | Yes | Yes |
FDA Form 2579 needs to be filed by the installer** | NA | NA | NA | NA | Yes | NA | NA | NA | NA |
*The post-move pre-examination test is described in the Policy Guidance Help System under “Quality Assurance/Equipment/Mobile Units Equipment Quality Control - Question #3.” This post-move pre-examination testing does not have to be performed by a medical physicist.
** The filing of Form 2579 is NOT an MQSA requirement. Under the Diagnostic X-Ray Performance Standards (21 CFR 1020.30(d)(1)), Form 2579 needs to be filed by the installer with the FDA.
Legend
- AB - Accreditation Body
- ACR - American College of Radiology
- AR - State of Arkansas
- CA - Certification Agency
- FDA - Food and Drug Administration
- IA - State of Iowa
- IL - State of Illinois
- NA - Not Applicable
- SC - State of South Carolina
- TX - State of Texas