U.S. flag An official website of the United States government

On Oct. 1, 2024, the FDA began implementing a reorganization impacting many parts of the agency. We are in the process of updating FDA.gov content to reflect these changes.

  1. Home
  2. Medical Devices
  3. Medical Device Safety
  4. Emergency Situations (Medical Devices)
  5. Coronavirus (COVID-19) and Medical Devices
  6. At-Home OTC COVID-19 Diagnostic Tests
  1. Coronavirus (COVID-19) and Medical Devices

At-Home OTC COVID-19 Diagnostic Tests

Español

Now available, U.S. households are eligible to order 4 free COVID-19 tests at www.COVIDTests.gov. The COVID-19 tests will detect current COVID-19 variants and can be used through the end of the year.

Many COVID-19 tests have extended expiration dates, so you may be able to use your COVID-19 tests after the expiration date that is printed on the box. To find out if your test has an extended expiration date, check the list of extended expiration dates.

On this page:


General Information About At-Home OTC COVID-19 Diagnostic Tests

A woman performing a nose swab on herself as part of a COVID at home test.

At-home over-the-counter (OTC) COVID-19 diagnostic tests can show if you have an active COVID-19 infection.

These at-home OTC COVID-19 diagnostic tests are FDA authorized for self-testing at home (or in other locations) without a prescription. Tests are available online or at local stores and  you collect your own sample, perform the test, and read the result yourself without the need to send a sample to a laboratory.

With most at-home OTC COVID-19 diagnostic tests, you should repeat testing following a negative result, whether you have symptoms or not, to reduce your risk of a false negative test result. The FDA encourages you to voluntarily and anonymously report your positive or negative test results every time you use an at-home COVID-19 test. You can send your test result to MakeMyTestCount.org or use an app or other digital option for self-reporting that may be included with your test. For additional information on reading and understanding your test results, see Understanding At-Home OTC COVID-19 Antigen Diagnostic Test Results.

For more information, including answers to frequently asked general questions about at-home COVID-19 diagnostic tests, visit At-Home COVID-19 Diagnostic Tests: Frequently Asked Questions.

Authorized At-Home OTC COVID-19 Diagnostic Tests and Expiration Dates

The table below is updated regularly and lists FDA-authorized at-home OTC COVID-19 diagnostic tests, including information on expiration dates, who can use the test, links to home use instructions for each test, and other details that may help you decide what test is right for you. For additional information about each Emergency Use Authorization (EUA), see In Vitro Diagnostics EUAs: Tables of IVD EUAs.

In the table below, the "Expiration Date" column lists where to find the expiration date for that test, and the "Other Details" column lists the shelf-life for the test. The shelf-life is how long the test should work as expected and is measured from the date the test was manufactured. The expiration date is set at the end of the shelf-life and is the date through which the test is expected to perform as accurately as when manufactured. In some cases, the expiration date for a test may be extended.

An extended expiration date means the manufacturer provided data showing that the shelf-life is longer than was known when the test was first authorized. For more information about how the expiration date is determined and why it may be extended, see the At-Home COVID-19 Diagnostic Tests: Frequently Asked Questions.   

To see if the expiration date for your at-home OTC COVID-19 test has been extended, first find the row in the below table that matches the manufacturer and test name shown on the box label of your test. If the expiration date has been extended, then the Expiration Date column states "Extended Expiration Date" and links to a document that lists the lot numbers, original expiration dates, and the extended expiration dates. If your lot number and/or original expiration date do not appear, or if the Expiration Date column states “Expiration Date: See box label,” do not use the test beyond the original expiration date on your test. 

Search the table below by manufacturer, test name, or other details shown in the table to find more information on the test, including if the expiration date has been extended. 

To see complete information on smaller screens, select the blue plus (+) button beside the test name.

Manufacturer and Test Name (Links to Instructions for Use)Expiration DateWho can use this test: SymptomsWho can use this test: AgeOther Details

Abbott Diagnostics Scarborough, Inc.:
BinaxNOW COVID-19 Ag Card Home Test

Packaging for Abbott Diagnostics Scarborough, Inc.: BinaxNOW COVID-19 Ag Card Home Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 15 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Requires supervision of a telehealth proctor and a smartphone or computer
  • Results in 15 minutes
  • Shelf Life: 22 months

Abbott Diagnostics Scarborough, Inc.:
BinaxNOW COVID-19 Antigen Self Test

Packaging for Abbott Diagnostics Scarborough, Inc.: BinaxNOW COVID-19 Antigen Self Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 15 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 22 months

Access Bio, Inc.:
CareStart COVID-19 Antigen Home Test

Packaging for Access Bio, Inc.: CareStart COVID-19 Antigen Home Test

Alternate brand name:

Packaging for On/go COVID-19 Antigen Self-Test
  • People with symptoms that began within the last 7 days
  • People without symptoms. The test is to be performed two times over three days (serial testing)
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Smartphone optional
  • Alternate brand name: On/go COVID-19 Antigen Self-Test
  • Results in 10 minutes
  • Shelf Life: 21 months

ACON Laboratories, Inc:
Flowflex COVID-19 Antigen Home Test

Packaging for ACON Laboratories, Inc: Flowflex COVID-19 Antigen Home Test

Alternate brand name:

Packaging for On/Go One COVID-19 Antigen Home Test
  • People with symptoms that began within the last 7 days
  • People without symptoms. The test is to be performed two times over three days (serial testing)
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Smartphone optional
  • Alternate brand name: On/Go One COVID-19 Antigen Home Test
  • Results in 15 minutes
  • Shelf Life: 24 months

Aptitude Medical Systems Inc.:
Metrix COVID-19 Test

Aptitude Medical Systems Inc.: Metrix COVID-19 Test package thumbnail
  • People with or without symptoms
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Molecular test
  • Nasal swab
  • Saliva
  • Requires a Metrix Reader (sold separately)
  • Smartphone optional
  • Results in 30 minutes
  • Shelf Life: 18 Months

CorDx, Inc.
CorDx TyFast Flu A/B & COVID-19 At Home Multiplex Rapid Test

Box for CorDx TyFast Flu A/B & COVID-19 At Home Multiplex Rapid Test
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 10 minutes
  • Shelf Life: 15 months

CTK Biotech, Inc.: 
ImmuView COVID-19 Antigen Home Test

ImmuView COVID-19 Antigen Home Test - Instructions for Use (Home Test)
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 20 months

Ellume Limited:
Ellume COVID-19 Home Test

Packaging for Ellume Limited: Ellume COVID-19 Home Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 4 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 16 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Mid-turbinate nasal swab
  • Requires a Smartphone
  • Results in 15 minutes
  • Shelf Life: 12 months

Genabio Diagnostics Inc.:
Genabio COVID-19 Rapid Self-Test Kit

Packaging for Genabio Diagnostics Inc.: Genabio COVID-19 Rapid Self-Test Kit
  • Expiration Date: See box label
  • People with symptoms that began within the last 7 days.
  • People without symptoms. The test is to be performed two times over three days (serial testing)
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 18 months

iHealth Labs, Inc.:
iHealth COVID-19 Antigen Rapid Test

Packaging for iHealth Labs, Inc.: iHealth COVID-19 Antigen Rapid Test

Alternate brand name:

Packaging for GoodToKnow COVID-19 Antigen Rapid Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 15 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Smartphone optional
  • Results in 15 minutes
  • Alternate brand name: GoodToKnow COVID-19 Antigen Rapid Test
  • Alternate brand name: GoToKnow COVID-19 Antigen Rapid Test
  • Shelf Life: 15 months

InBios International Inc:
SCoV-2 Ag Detect Rapid Self-Test

Packaging for InBios International Inc: SCoV-2 Ag Detect Rapid Self-Test
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 20 minutes
  • Shelf Life: 13 months

Immunostics Inc.: Swab-N-Go Home Test COVID-19 Ag 

Box Label:  Immunostics Inc.: Swab-N-Go Home Test COVID-19 Ag
  • Expiration Date: See box label 
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes 
  • Shelf Life: 20 months 

Pfizer Inc:
Lucira CHECK-IT COVID-19 Test Kit

Pfizer Check Box Label
  • People with or without symptoms
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Molecular test
  • Nasal swab
  • Smartphone optional
  • Results in 30 minutes
  • Shelf Life: 18 months

Maxim Biomedical, Inc.:
MaximBio ClearDetect COVID-19 Antigen Home Test

Packaging for Maxim Biomedical, Inc.: MaximBio ClearDetect COVID-19 Antigen Home Test
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 24 months

Mologic, Inc.:
COVI-Go SARS-CoV-2 Ag Self-Test

Mologic, Inc.: COVI-Go SARS-CoV-2 Ag Self-Test Box Label
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 20 minutes
  • Shelf Life: 6 months

OraSure Technologies, Inc.:
InteliSwab COVID-19 Rapid Test

Packaging for OraSure Technologies, Inc.: InteliSwab COVID-19 Rapid Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 18 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Smartphone optional
  • Results in 30 minutes
  • Shelf Life: 24 months

OSANG LLC:
OHC COVID-19 Antigen Self Test

OHC COVID-19 Antigen Self Test Packaging

Alternate brand name:

Quickfinder Covid-19 Antigen Self Test Packaging
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing)
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Alternate brand name: QuickFinder COVID-19 Antigen Self Test
  • Results in 15 minutes
  • Shelf Life: 24 months

OSANG LLC:
QuickFinder COVID-19/Flu Antigen Self Test

QuickFinder COVID-19/Flu Antigen Self Test Packaging
  • See box label
  • People with symptoms that began within the last 4 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 18 months

PHASE Scientific International, Ltd.:
INDICAID COVID-19 Rapid Antigen At-Home Test

Packaging for PHASE Scientific International, Ltd.: INDICAID COVID-19 Rapid Antigen At-Home Test
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 20 minutes
  • Shelf Life: 15 months

Quidel Corporation:
QuickVue At-Home OTC COVID-19 Test

Packaging for Quidel Corporation: QuickVue At-Home OTC COVID-19 Test

Alternative Brand Names:

CVS Health At Home COVID-19 Test Kit

 

Walgreens At-Home COVID-19 Test Kit
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Alternative Brand Names: CVS Health At Home COVID-19 Test Kit and Walgreens At-Home COVID-19 Test Ket
  • Results in 10 minutes
  • Shelf Life: 16 months

Watmind USA: Speedy Swab Rapid COVID-19 + Flu A&B Antigen Self-Test

Box Label of the Speedy Swab Rapid COVID-19 + Flu A&B Antigen Self-Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 18 months

SD Biosensor, Inc.:
Pilot COVID-19 At-Home Test 1

Packaging for SD Biosensor, Inc.: Pilot COVID-19 At-Home Test

FDA Safety Communication - Recall of Certain Lots

  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 20 minutes
  • Telehealth Proctor Supervised (optional)
  • Shelf Life: 16 months

Siemens Healthineers:
CLINITEST Rapid COVID-19 Antigen Self-Test

Packaging for Siemens Healthineer: CLINITEST Rapid COVID-19 Antigen Self-Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Smartphone/Web access (optional)
  • Shelf Life: 24 months

Watmind USA:
Speedy Swab Rapid COVID-19 Antigen Self-Test

Packaging for Watmind USA: Speedy Swab Rapid COVID-19 Antigen Self-Test
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing)
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 21 months

Xiamen Boson Biotech Co., Ltd.:
Rapid SARS-CoV-2 Antigen Test Card

Packaging for Xiamen Boson Biotech Co., Ltd.: Rapid SARS-CoV-2 Antigen Test Card
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 18 months

SEKISUI Diagnostics, LLC.:
OSOM COVID-19 Antigen Home Test

OSOM COVID-19 Antigen Home Test - box label
  • Expiration Date: See box label
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 11 months

Beijing Hotgen Biotech Co., Ltd.:
Hotgen COVID-19 Antigen Home Test

Packaging for Hotgen COVID-19 Antigen Home Test
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 10 months

CorDX, Inc.:
CorDx COVID-19 Ag Test

Packaging for Axteria BioMed Consulting Inc.: CorDx COVID-19 Ag Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 10-30 minutes
  • Shelf Life: 15 months

Azure Biotech Inc.:
Fastep COVID-19 Antigen Home Test

Packaging for Azure Biotech Inc.: Fastep COVID-19 Antigen Home Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 24 months

Oceanit Foundry LLC:
ASSURE-100 Rapid COVID-19

Photo of Oceanit Assure Covid Test Box
  • Expiration Date: See box label
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 20-30 minutes
  • Shelf Life: 12 months

Advin Biotech Inc.:
Advin COVID-19 Antigen Test @Home

Photo of AdvinBio_Covid16 Test box
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life:  24 months

GenBody Inc.:
GenBody COVID-19 Ag Home Test

Image of the GenBody COVID-19 Ag Home Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-20 minutes
  • Shelf Life: 24 months

Azure Biotech, Inc.:
Azure FaStep® COVID-19 Antigen Pen Home Test

Azure FaStep® COVID-19 Antigen Pen Home Test Box Label
  • Expiration Date: See box label
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 21 months

Pfizer Inc.:
Lucira by Pfizer COVID-19 & Flu Home Test

Lucira COVID-19 & Flu Home Test - Box Label
  • People with symptoms
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Molecular test
  • Nasal swab
  • Results in 30 minutes
  • Shelf Life: 18 months

Princeton BioMeditech Corp:
Status COVID-19 Antigen Rapid Test for Home Use

Princeton Status COVID-19 Antigen
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 24 months

Nano-Ditech Corporation:
Nano-Check COVID-19 Antigen At-Home Test

Packaging illustration
  • Expiration Date: See box label 
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 15 minutes 
  • Shelf Life: 6 months 

BioTeke USA, LLC:
Bio-Self COVID-19 Antigen Home Test

Packaging for Bio-Self COVID-19 Antigen Home Test
  • Expiration Date: See box label
  • People with symptoms that began within the last 7 days. The test is to be performed two times over three days (serial testing).
  • People without symptoms. The test is to be performed three times over five days (serial testing)
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15-30 minutes
  • Shelf Life: 12 months

3EO Health, Inc.:
3EO Health COVID-19 Test

Packaging for 3EO Health COVID-19 Test
  • Expiration Date: See box label
  • People with symptoms.
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Molecular test
  • Nasal swab
  • Results in 30 minutes
  • Shelf Life: 4 months

SEKISUI Diagnostics, LLC:
OSOM Flu SARS-CoV-2 Combo Home Test

Packaging for SEKISUI Diagnostics, LLC: OSOM Flu SARS-CoV-2 Combo Home Test
  • People with symptoms that began within the last 4 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 10-30 minutes
  • Shelf Life: 12 months

Wondfo USA Co., Ltd.
WELLlife COVID-19 / Influenza A&B Home Test

EUA-WELLLife Flu COVID OTC Box Label
  • Expiration Date: See box label
  • People with symptoms that began within the last 5 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 10 minutes
  • Shelf Life: 6 months

iHealth Labs, Inc.:
iHealth COVID-19/Flu A&B Rapid Test

iHealth Labs, Inc.: iHealth COVID-19/Flu A&B Rapid Test Product labeling
  • People with symptoms that began within the last 4 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 17 months

ACON Laboratories, Inc.:
Flowflex Plus COVID-19 and Flu A/B Home Test

Flowflex Plus COVID-19 and Flu A/B Home Test - Image of Packaging
  • Expiration Date: See box label
  • People with symptoms that began within the last 6 days. The test is to be performed two times over three days (serial testing).
  • Age 14 years and older
  • Age 2 years and older when collected by an adult 
  • Antigen test
  • Nasal swab
  • Results in 15 minutes
  • Shelf Life: 15 months

Related Information:

  • 1The Pilot COVID-19 At-Home Test, distributed by Roche Diagnostics, is previously known as the COVID-19 At-Home Test. As the manufacturer, SD Biosensor, transitions to this new brand, customers may receive a product with the previously authorized COVID-19 At-Home Test labeling.

Subscribe to Consumer Information for Medical Devices

Receive notifications for consumers about medical device information, recently approved devices, and alerts that may be of interest to the general public.

Back to Top